Унискан 2000
- minimal radiation dose of patient and medical staff;
- high quality and diagnostic value of digital image;
- low operating costs;
- availability of operation in platforms.
Universal digital X-ray system is meant for X-ray examination of patients in the standing position, in the sitting position and in the lying position.
Principle of operation: digital projection image is formed by linear scan of patient with fan-shaped X-radiation beam with synchronized detection of absorbed radiation by the instrumentality of multiple-unit semiconductor detector (matrix).
Advantages:
![]() |
![]() |
![]() | |
![]() |
![]() |
||
![]() |
![]() |
- high-frequency X-ray generator;
- X-ray tube with flat-field collimator;
- lifting-turning support device;
- radiotransparent table for patient;
- digital scanning receiver of X-radiation;
- X-ray dosimeter for measuring dosimetric cost on patient;
- computer work stations of radiologist and laboratory assistant;
- power supply.
Composition:
Manufactured jointly with НП ЧУП "Адани" Minsk | ||
Technical characteristics | ||
---|---|---|
High-frequency X-ray generator | ||
Power | 1,5 kW | |
Range | 50-125 kW with step of 1 kW | |
Instability for 8 s | ±0,5% | |
Pulsation rate | less than 0,2% | |
Frequency | 25 kHz | |
Maximal anode current | 12,5 mA at 125 kW; 20 mA at 50 kW | |
Exposure time | up to 8 s | |
Power supply | 380V/50Hz | |
X-ray tube | 12°, 0,3/1.0 mm, 3,4/35,5 kW, 150 kW, filter 4,5 mm AI | |
Digital scanning receiver | ||
Detector type | 1152-element digital matrix of silicon photodiodes with Csi (TI) coating | |
Image ration | 1152 pixels x 1152 pixels x 16 bit | |
Contrast resolution | not worse than 1% | |
Imaging time | 8 s | |
Equivalent dose for patient per examination | not more than 25mcSv | |
Work station of radiologist | Pentium II-400/32Mb RAM/HDD 4,5Gb/Video 8Mb/Monitor 15" SVGA/UPS 300W |
|
Work station of laboratory assistant | Pentium II-400/64MB RAM/HDD 8Gb/Video 16Mb/Monitor 19" SVGA/UPS 600W/Printer HP LaserJet 2100/MO drive 3,5" with capacity of 640Mb/10 MO disks for archiving |
|
X-ray dosimeter | Ionization chamber 250 mm x 25 mm on collimator output |