Comparison microscope «Пеленг МС-4»
- analysis of objects in reflected radiation, ultra violet radiation, visible radiation, infrared radiation, polarized visible radiation and in transmitted (against the light) visible radiation;
- visual observation of images of examined object through eye-pieces of binocular head or on the screen of computer;
- recording images of analyzed object on digital camera;
- synchronized change of magnification in both channels;
- evaluation measurement of linear dimensions between elements of analyzed objects.
Comparison microscope is meant for identification of objects or traces, scratches, imprints etc. on various objects when carrying out different forensic research in expert sub-units of security organizations, as well as with scientific purpose.
Manufactured jointly with ОАО "Пеленг"Minsk | |
---|---|
Technical characteristics | Пеленг МС-4 |
Magnification at visual observation, zoom, with error of not more than 5% | 5; 10; 20; 40; 80 |
Viewing field of microscope depending on magnification, mm, not less than | 40; 20; 10; 6; 2,5 |
Output signal of built-in video camera | Standard, television (1 V, 75 Ohm) |
Focusing range, mm, not less than | 15 |
Distance between lens and front plane of the table, mm | 115±10 |
Shift of object plane of microscope stages: - linear in horizontal plane along two mutually perpendicular directions from middle position, mm - rotatory of horizontal plane, degree - linear motorized shift along vertical, mm - mutually motorized shift of two microscope stages in horizontal plane along the line connecting axes of two channels from middle position in both directions, mm |
±30 360 240±10 100±5 |
Range of eye-piece diopter focusing divergence, D | from +5 up to -5 |
Range of interpupillary distance change, mm | from 56+10 up to 75-1,0 |
Range of diameters of analyzed cartridge cases and bullets, mm | 3 - 25 |
Range of lengths of analyzed cartridge cases and bullets, mm | 4 - 70 |
Range of evaluation measurements of linear dimensions, mm | 0 - 60 |
Minimal division value of scale rule, mm | 0,05 |
Inaccuracy of measurements of linear dimensions, mm | 0,05 |
Microscope supply voltage of alternating current of frequency(50±1)Hz, V | 220±22 |
Rated power consumption, V·A, not more than | 650 |
Speed of motorized shift of tables, mm/s not more than | 5 |
Distance between floor and microscope eye-pieces, mm, not more than | 1150 |
Overall dimensions, mm, not more than | 800х600х1250 |
Weight, kg, not more than | 160 |